3. Locate the write protection switch. It is the second switch (2) marked P.
4. To write protect the level gauge, move switch P to the upper position.
5. Check that contact surfaces on the housing and cover are clean. Replace the cover
and tighten the screws. Make sure that the cover is fully engaged to meet
explosion-proof requirements and to prevent water from entering the terminal
compartment.
Note
Ensure that O-rings and seats are in good condition prior to mounting the cover in
order to maintain the specified level of ingress protection.
6.2.8
Logging measurement data using TankMaster
™
The Rosemount 5900S supports logging of diagnostic registers. This function is useful for
verifying that the gauge works properly. The logging function can be accessed by using
the Rosemount TankMaster WinSetup program.
Procedure
1. Start the Rosemount TankMaster WinSetup program.
2. In the
TankMaster WinSetup
workspace window, select the device icon.
Reference Manual
Service and troubleshooting
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
199
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...