Free space
The following free space is recommended for mounting the Rosemount 5900S with
LPG/LNG Antenna:
Figure 3-15: Free Space Requirements for Rosemount 5900S with LPG/LNG antenna
C
A
B
D
E
A. Recommended space 550 mm (22 in.) for installation and service
B. Recommended space 1000 mm (39 in.) for installation and service
C. Minimum 1200 mm (47 in.) to product surface for highest accuracy. Minimum 800 mm
(31 in.) with reduced accuracy
D. Optional pressure transmitter
E. Product surface
Reference Manual
Installation
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
45
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...