Backing up multiple device configurations using
TankMaster
™
Using Rosemount TankMaster WinSetup to save configuration for multiple devices:
Procedure
1. Start the Rosemount TankMaster WinSetup program.
2. In the
WinSetup
workspace window select the
Devices
folder.
3. Click the right mouse button and choose the
Save Database of All to Files
option.
This option is also available from the
Service/Devices
menu.
4. Select a device from the
Available Devices
pane and press the
Add
button in order to
move it to the
Selected Devices
pane. Repeat for all devices you wish to include.
5. Choose the
Holding Registers
and
Predefined Registers
options (the User-Defined
option should only be used for advanced service).
6. Click the
Browse
button, select a folder and type a name for the backup file.
7. Click the
Start
button to save the database backup.
The backup file can be viewed as a text file in any word processing program:
Reference Manual
Service and troubleshooting
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
191
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...