Table 6-10: RECOMMENDED_ACTION
Alert Type
Host Diagnostic
Message
Description
Recommended Action
None
N/A
None
No action required
Failure
Software
Incompatibility
Error
FF I/O Board software and radar level
gauge main firmware versions are
incompatible.
The device is not in service (OOS).
1. Replace the transmitter head.
2. Contact Emerson Automation
Solutions/Rosemount Tank
Gauging department.
Memory Failure -
FF I/O Board
Configuration data has been corrupted
or pending configuration changes has
been lost due to power loss before
data storage was completed.
Default values are loaded into the
faulty block. Potential errors in stored
data may cause unwanted behavior.
The device is not in service (OOS) and
status for all variables is BAD. Device
recovery is possible.
1. Do Factory Reset - FF I/O Board.
2. If error persists, it may indicate a
faulty memory chip. Replace the
transmitter head.
Device Error
The transmitter head is out of order.
Field repair might be possible.
The device is not in service (OOS).
No valid measurement values can be
retrieved.
1. Replace the transmitter head.
Internal
Communication
Failure
The communication between the
radar level gauge main board and the
FF I/O Board has been lost
1. Replace the transmitter head.
Electronics Failure The device has detected a fault with an
electrical component on the FF I/O
Board module assembly.
The device is not in service (OOS).
1. Replace the transmitter head.
Out of
Specification
Device Major
Information
Measurement values are retrieved but
device needs to be serviced.
Problem with installation or physical
environment that might affect
measurement and device behavior in
the long term.
See Device Status for more
information on the error source (see
Viewing device status in AMS Device
Manager
).
1. Check the mechanical installation
and environment.
Service and troubleshooting
Reference Manual
September 2019
00809-0100-5900
218
Reference Manual
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...