Table 3-13: Terminal Block Connections for the Rosemount 5900S with SIL Option
Connection
SIL Safety System
X1: Primary Tankbus in Intrinsically safe Tankbus input, power and communication
X2: Alarm
SIL3 alarm output
(connect to Exi terminal block on Rosemount 2410 Tank Hub)
X3: Primary Tankbus
out
Optional jumpers between X3 and X4 for connection to second level
gauge of the Rosemount 5900S / 2-in-1 version
X4: Secondary Tankbus
in
Test terminals
Test terminals for temporary connection of a field communicator
See the Rosemount 5900 Radar Level Gauge and 2410 Tank hub
for Use in
Safety Instrumented Systems for more information on how to setup a Rosemount Tank
Gauging SIL3 safety system.
3.4.12
Wiring diagrams
The standard version of the Rosemount 5900S has a single intrinsically safe fieldbus input.
You may activate a built-in fieldbus termination by short-circuiting the X2 connector.
An intrinsically safe output on connector X4 can be used for “daisy-chain” connection to
other devices in a Rosemount Tank Gauging system.
Connector X3 is used for a fieldbus input/output cable shield connection (separated from
chassis ground).
illustrates a typical wiring diagram with a Rosemount 5900S level gauge
connected to a Rosemount 2240S temperature transmitter. In this example the
termination is enabled in the temperature transmitter which is the last device on the
Tankbus (see
).
In case you prefer to connect the temperature transmitter to the tank hub, you may
“daisy-chain” the Rosemount 5900S to the temperature transmitter, and terminate the
Tankbus by a jumper in terminal X2 on the Rosemount 5900S terminal block.
Installation
Reference Manual
September 2019
00809-0100-5900
90
Reference Manual
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...