17. Select the
Gas Correction
tab.
18. Choose the appropriate correction method for the product in the tank:
Option
Description
Air Correction
This method should only be used when there is no vapor in
the tank, i.e. when the tank is empty and contains air only. It
is used in the initial step when calibrating the Rosemount
5900S.
One known gas
This method may be used when there is only one gas type in
the tank. It provides the highest accuracy among the
different correction methods. Note that even small amounts
of another gas reduces the accuracy.
One or more
unknown gases
Use this method for hydrocarbons, for example Propane/
Buthane, when the exact mixture is not known.
Two gases with
unknown mixratio
This method is suitable for a mixture of two gases even if the
mixratio is not known.
One or more known
gases with known
mixratio
This method may be used when there is a well known
mixture of up to 4 products in the tank.
Now the Rosemount 5900S level gauge is ready to measure the product level when
the tank is put into operation.
F
OUNDATION
Fieldbus parameters:
TRANSDUCER 1500>LPG_CORRECTION_METHOD
TRANSDUCER 1500>LPG_NUMBER_OF_GASSES
TRANSDUCER 1500>LPG_GAS_TYPE1, TRANSDUCER 1500>LPG_GAS_PERC1
Configuration
Reference Manual
September 2019
00809-0100-5900
180
Reference Manual
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...