6.2.3
Recover a backup configuration database using
TankMaster
™
Rosemount TankMaster WinSetup lets you replace the current Holding Register database
with a backup database stored on disk. This can be useful, for example, if you want to
recover lost configuration data.
To load a Holding Register database do the following:
Procedure
1. In the
TankMaster WinSetup
workspace window, select the device icon.
2. Right-click and select
Upload Database
, or from the
Service
menu choose
Devices/
Upload Database
.
3. Click the
Browse
button and choose a database file to be uploaded, or type a path
and file name.
4. Click the
Upload
button.
6.2.4
View and configure diagnostic registers using
TankMaster
™
The Rosemount TankMaster WinSetup program lets you view the current device status.
The
View Diagnostic Register
window shows a selection of database registers that gives
Service and troubleshooting
Reference Manual
September 2019
00809-0100-5900
192
Reference Manual
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...