Adjustments
92
Purge pressure
Fig. 72: PR2
PR2 is the pressure regulator for the purging system of the IR-cell and it is situated inside the
analyzer, just behind the left-side of the panel. It is positioned horizontally.
–
The oxygen pressure at the outlet of the PR2 should be 0.35 bar (5psi). The pressure can
be checked by applying a T-piece on the outlet of the pressure regulator, and connecting
a pressure gauge (1), as indicated above.
–
Pull and rotate the knob of the pressure regulator (PR2), until the correct
pressure is
shown on the gauge.
Pos : 5. 14 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Cl osing c one adjust ment ELTR A @ 6 \ mod_1369381713916_9. doc x @ 42030 @ 2 @ 1
4.7 Closing cone adjustment
Pos : 5. 15 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 10Adjust ment s_C S-2000/ 1035 M odul Cl osing cone @ 9\mod_1425542547459_9. doc x @ 74634 @ @ 1
–
Ensure that the shaft (1) is securely fastened to the (1). (engineers wrench size 17).
–
In order to adjust the cone, the screws (4) should be loosened so that the cone (3) moves
horizontally on the bar (2).
–
Close the furnace ( with the lift upwards ).
–
Tighten the screws (4).
–
The cone is now positioned correctly.
–
The screws (5), (6), (7), must now be well tightened in this order.
–
The screws (8) should be properly fastened (with a wrench).
–
The screws (9) should be tightened by hand only.
bar
0.5
1
1.5
2
2.5
0
P R 2
1
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...