Adjustments
84
Pos : 5. 8 / 0025 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 Infrar ed cell temper at ure regul ati on ELTRA @ 6\ mod_1369641998919_9. doc x @ 42289 @ 2 @ 1
4.4 Infrared cell temperature regulation
Pos : 5. 9 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 2000_s er vic e/10Adj ust ments _CS- 2000/1020 Modul IR ther mos tatic c ontrol @ 9\ mod_1425542463669_9.doc x @ 74598 @ @ 1
IR cell temperature control
Fig. 65: IR cell temperature control
TP28 is the test point for measuring the set point of the IR temperature against TP1=GND.
At 3000 mV the temperature is 27°C growing with 10 mv /°C.
Examples:
3.10V
37°C
3.30V
57°C
This is the adjustable temperature range of the cell in normal cases. This temperature range is
adjustable when there is a jumper set on position 6-Lo temperature of JP1 and mainly there is
no jumper on position 5-Hi Temp of JP1. In other words, when no jumper is set, this
temperature range is already by default. A jumper in position 6-Lo Temp will keep this range, so
there is no difference, but the jumper will be available when a jumper is elsewhere really
needed.
3.73V
100°C
3.98V
125°C
This temperature range is available for heated paths only. They are the paths for measuring
water resp. hydrogen in the analyzers CW and CHS and possibly later in CHNS analyzers. In
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...