Faults
32
Pos : 3. 12 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Anal ys edauer zu lang - The anal ysis t akes t oo long ELTR A @ 6\ mod_1368613082100_9.doc x @ 40314 @ 2 @ 1
3.6 Analysis takes too long
Pos : 3. 13 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0535 M odul T he anal ysis takes too l ong @ 9\ mod_1425539613183_9. doc x @ 74165 @ 3444 @ 1
Fig. 24: Analysis time
The analysis respectively the combustion, takes normally a maximum of 100 seconds. At the
beginning of combustion, the IR-cells signals start going up on the computer screen, reaching a
peak maximum depending on the C resp. S concentrations in the sample. When the signal
comes down, the analysis will be terminated when the signal has reached a comparator level.
The value of this comparator level is entered in the configuration window of the software.
If, for some reason, the signal may not come down to the comparator level, the analysis is
terminated when the maximum analysis time expires. This time is also entered in the
configuration window.
The analysis stops also when the maximum analysis time expires before the signal comes down
to the comparator level.
The meaning of setting a maximum analysis time is to protect the induction generator from
overheating, in case the signal may not come down to the comparator level. One reason why
the signal not to come down to the comparator level is that Hydrocarbons from the oxygen burn
on the hot crucible, preventing the IR-signal from coming down to zero resp. down to the
comparator level. One more reason is a too low comparator level, so that a small positive drift of
the base line during analysis makes the base line higher than the comparator level at the end of
the combustion. In such cases, the comparator level should be set higher.
The comparator level value entered into the configuration window is a minimum value which
increases according to the peak maximum of each analysis. During combustion, the software
takes the peak maximum to add to the entered comparator value an additional amount
depending on the actual peak maximum.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...