Service
102
NOTICE
The position of the IR sources in their housing has to be as shown in the graphics A and B.
Correct:
Wrong:
Fig. 86: IR sources in their housing
If the position of a source is wrong, the radiation reaching the IR sensor is lower than when the
source is in the right position. It may happen that after transportation due to a strong shock that
the base lines are not in the right level due to the sources being not in the right position any
more. For correct base
lines see chapter “
Infrared cell adjustments
”.
(Take care in case of previous IR cell models having a base line around zero volts.)
For correcting the position of bent IR sources use a thin forceps or carefully push them to the
right central position by using a thin screw driver. Take care not to bend the wires too far. They
are quite hard so that when bending back and forth they may break.
In case of changing a source, the wires must be very properly soldered because the resistance
of a source is below one Ohm.
All IR sources for all available cells in a module are connected in series.
For adjusting the IR source voltage see chapter
“
IR-source voltage setting
”.
Pos : 7. 6 / 0025 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 C hopper ELTR A @ 6\ mod_1369641924960_9. doc x @ 42267 @ 2 @ 1
5.3 Chopper
Pos : 7. 7 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 2000_s er vic e/15 Ser vice_CS- 2000/ 1520 Modul C hopper @ 9\ mod_1425543409739_9.doc x @ 74742 @ @ 1
Motor replacement
:
The replacement motors are supplied with their wires already soldered on a 9-pin plug, so that
in case of replacing a chopper motor, only the two wires of the IR sources have to be soldered
on the connector.
C
D
E
F
5
5
5
5
B
A
5
6
5
3 - 2 - 1
C
D
E
F
5
5
5
5
B
A
5
6
5
3 - 2 - 1
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...