Faults
66
analysis, the burned sample in the crucible looks like normally combusted sample. If so, check
the bypass valve V4. See chapter
“
Cleaning Solenoid valves
”.
Pos : 3. 68 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 IR signal does not c ome down t o the base line at t he end of anal ysis @ 7\ mod_1381404509510_9. doc x @ 49850 @ 2 @ 1
3.34 IR signal does not come down to the base line at the end of analysis
Pos : 3. 69 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0685 M odul IR signal does not c ome down t o the base line @ 9 \ mod_1425541763229_9. doc x @ 74537 @ 3333 @ 1
There are several different reasons for this effect
1.
Long combustion of a very low carbon sample:
Fig. 51: Programm screenshot (graph example 1)
It appears like a long tailing at the end of the peak. It appears in the low carbon range only and
the distance to the base line is in the range of 10 to 50mV. Depending on the blanks of the
crucibles and on the oxygen purity, the level of the tailing can be as high as 100mV. When the
burning is extremely long, the level of the low carbon tailing even starts to grow.
Possible solutions:
-
Reduce the analysis time
-
Increase the comparator level of the low carbon range,
-
Reduce the generator time.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...