Adjustments
85
order to have the temperature adjustable in the above range, a jumper has to be set on position
5-Hi Temp. of JP1.
The temperature is set by turning the pot and measuring the set value on
The actual value of the temperature is measured on TP26.
Caution: After changing the temperature set point, it takes some time until the actual
temperature value is stabilized at the new set level.
Fig. 66: Trimmer
This is the trimmer in the IR cell module for adjusting the cell temperature. It is accessible with a
screw driver from outside the cell. There is no need to remove the cell block module from the
analyzer’s cabinet in order to adjust the temperature
Modifications for higher sensitivity (gain) in the low sulphur range:
For LS: R101=47k R100 = 3M
4.1 x Standard gain
Recalculation factor = 92.664
For double the normal sensitivity: R101=180k
1.92 x standard gain
Recalculation factor = 39.5
For standard gain.
Recalculation factor = 20.59
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...