Faults
52
When the pulses at TP13 are present without starting analysis and with 12 V DC at TP9, then
the circuit board HF 42 is out of order.
If no scope is available, the pulses can be detected by using a voltmeter.
Normal values are about 0 V AC without pulses, and about 1 V AC with pulses.
This indicated value can be different, depending on the type of the voltmeter used.
If this AC voltage is practically zero but even though the generator is powered, the next possible
reason: The triac is faulty (short circuit), and has to be replaced.
Fig. 40: Circuit board HF 42
Pos : 3. 40 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Cr ucibl e or boat cr ac ks or melts ELTRA @ 6 \mod_1368708007091_9. doc x @ 41231 @ 2 @ 1
3.20 Crucible or boat cracks or melts
Pos : 3. 41 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0615 M odul Cr ucibl e cr ac ks or melts @ 9\ mod_1425540955461_9.doc x @ 74357 @ 33 @ 1
3.20.1 Induction furnace
One reason could be that the quality of the crucible is bad. Use a better quality.
Another reason could be that the furnace temperature is too high, due to:
–
Excessive main voltage.
–
Excessive sample and / or accelerator weight.
–
The use of a "sample or accelerator" that requires lower generator power.
In this case reduce the generator power in case of induction furnace. See
chapter „
Induction
generator control
“.
3.20.2 Resitance furnace
In case of resistance furnace, the boat melts when there is excessive amount of metal in the
combustion boat. The dangerous thing is that the molted metal destroys not only the boat but
also the combustion tube, finally dropping thru the tube on the thermocouple. It is then very hard
to remove the deformed tube from the furnace even after breaking it in two.
Pos : 3. 42 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 O xygen pres sur e goes up ver y sl owl y @ 9 \ mod_1425548475420_9. doc x @ 74976 @ 2 @ 1
3.21 Oxygen pressure goes up very slowly
Pos : 3. 43 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0620Modul T he oxyg en pr ess ure g oes up ver y slowl y @ 9\ mod_1425540999219_9. doc x @ 74369 @ @ 1
When using induction furnace operation:
After the button has been pressed to close the furnace (or to start the analysis by closing the
furnace first ), the oxygen gauge (6) should reach the normal level of
1,5 bar within a couple of seconds after the furnace has closed.
If this takes longer:
Then the problem is most probably outside the analyser.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...