Faults
46
After starting an analysis”, the following message appears: "
no oxygen pressure
".
If this message permanently remains in the “status window”, then check
–
Is the power switch (1) set on position 3?
–
The pressure on the oxygen gauge (2) may be too low. 1.5bar is normal
(22psi). The
pressure remains constant when the furnace is closed, and it sinks during the upward
movement of the furnace floor.
–
Check the oxygen pressure in the supply line or tank. 4 to 6 bar (60 to 90psi)
showing on the pressure gauge is normal.
–
Check if the pressure gauge for compressed air (5) shows 4 to 6 bar (60 to 90 psi).
–
Check if the cock (3) is properly shut, or else the O-ring (4) will not seal
Check if the
O-ring (4) is in a good condition.
–
Check if the combustion tube is broken or cracked.
–
Do leakage check. See chapter: “
Leak checking
”.
–
Check if the chemicals are saturated and have become hard.
–
The paper filter could also be the cause. For changing the paper filter, read the
operating manual.
–
See chapter
“
Fluctuating gas flow
”
– The contents of the gas cleaning furnace could be worn out and clogged.
Oxygen solenoid valve (V5) may be defective:
–
Open the left-side panel of the analyzer.
–
Unscrew the upper tube (1). The oxygen should rush out when the power
switch is set
to pos. 3
If not:
–
There may be no oxygen supplied.
Fig. 33: Terminal of the solenoid coil
–
220V AC is not measured on the terminal (2) of the solenoid coil.
–
The solenoid valve is clogged.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...