Faults
67
2.
Copper oxide has a bad quality
:
Fig. 52: Programm screenshot (graph example 2)
Memory effects of the copper oxide can cause tailing of the low carbon range in the level of 10
to 50 mV. However the tailing level does not grow due to faulty copper oxide.
Possible solution:
-
Replace the copper oxide
3.
Path window is not sealed or it is broken
If the faulty window is on the chopper housing side, all available cells have a tailing due to
combustion gases entering the chopper housing causing permanent IR absorption for all cells
available.
If the faulty window is on the IR sensor side, there is tailing of the damaged cell only.
Solution:
-Replace the part having the faulty window.
Replacing the window only, is rather tricky because the window should be available, the faulty
window must be removed, the surface for gluing new window must be cleaned from old glue
and rests of the broken window, there must be glue available being suitable to glue windows
(two components glue) and time should also be available to wait until the glue is solid.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...