Faults
39
3.10.2 Other faults
–
The screw for the upper furnace cushioning is too tight, see chapter
“
Pneumatics
”
–
The pressure switch does not close. Measure the voltage across its leads. Zero volts with
pressure, 24 VDC without pressure.
–
The furnace closes too slowly, adjust according to chapter
“
Pneumatics
”
–
The furnace does not close: see chapter
“
The induction furnace does not close
”
3.10.3 Resistance furnace operation
–
Check the oxygen pressure on the upper pressure gauge. The pressure should be about
1,5 bar (22 psi). If lower, see
Chapter “
No or low oxygen pressure
”
–
Check the pump voltage in the “Device state window”.
Pos : 3. 22 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 IR-c ell t emper at ure c ontrol is out of r ange ELTR A @ 6\ mod_1368687694553_9. doc x @ 40448 @ 2 @ 1
3.11 IR-cell temperature control is out of range
Pos : 3. 23 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0565 M odul IR t emperatur e c ontrol out of range @ 9\ mod_1425539916703_9.doc x @ 74237 @ 3 @ 1
If in the device state the graphs of IR-temperature °C or %IR are not in the green area, adjust
the temperature. See chapter
“
IR-cell temperature control
”.
Fig. 27: Device state
–
Thermostatic control reports 0% or 100% power:
The percentage of power required by the thermostatic control is displayed in the Device State
Window
. The power is 100% after switching on from cold while heating up. It takes about
30 minutes for the temperature to reach the set point. In stable condition, the power required
should be between 30 and 70%.
If the value is less than 30%, adjust a higher temperature set point. If the value is over 70%
without any tendency for coming down to lower values, adjust a lower set point.
Pos : 3. 24 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Breaker t urns of f at the begi nni ng of c ombusti on ELTR A @ 6 \ mod_1368688648887_9. doc x @ 40470 @ 2 @ 1
3.12 Breaker turns off at the beginning of combustion
Pos : 3. 25 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0570 M odul Breaker t urns of f at th beginning of c umbustion @ 9\ mod_1425539992083_9.doc x @ 74249 @ @ 1
If the circuit breaker turns off at the beginning of combustion then:
–
Unplug the analyzer mains power plug.
–
Switch on the breaker.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...