Faults
27
One reason could be the resistor (8) having a short circuit. Normally its resistance has the value
of 200K.
One more possibility is a faulty capacitor. Check the capacitors, mainly the one on top (7) of the
oscillator circuit which is connecting the anode of the tube with the oscillator circuit. Check also
the grid capacitor (6) which is placed underneath the oscillator chassis. Try to find out whether
they are broken by applying a reasonable force on them to see whether they are still in one
piece.
Check also the other, bigger capacitors (9), (10) and (11) however if one of them was faulty,
most probably the fuses would blow.
3.3.2 Resistance furnace operation
See Chapter 2.4.2
“IResitance furnace operation”
.
Pos : 3. 8 / 0025 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 Bad combusti on ELTRA @ 6\ mod_1368608656440_9. doc x @ 40248 @ 2 @ 1
3.4 Bad combustion
Pos : 3. 9 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 2000_s er vic e/05F aul ts_C S-2000/0520 M odul Bad combusti on @ 9\ mod_1425539443621_9. doc x @ 74129 @ 33 @ 1
3.4.1 Induction furnace operation
In most cases the reasons for bad combustion are:
–
Problems with the flow
–
The Generator (oscillator) tube is not powerful enough anymore. Norma
lly this does
not occur suddenly. It is rather a long process.
(Part number 77210)
–
The contacts of the tube are bad.
Some more hints:
Fig. 19: bypass valves
The bypass valve V4 (28) is either falsely activated during combustion, or the plunger jams. In
this case, the oxygen flows through the bypass valve into the IR-cell and not through the
furnace. The furnace remains nonetheless under oxygen pressure.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...