Faults
13
Notice
The IR cells have normally base lines adjusted to about 8V and the peak comes down during
analysis, respectively during IR absorption by the measuring gas.
In case of golden paths, in the first couple of months after being in operation, the base line may
goes up, which means that the paths become more reflective during this period. Therefore, the
base lines of golden paths should be initially set rather at 5 or 6 volts instead of 8 volts. In the
first couple of months the base lines go up, but the growing slows down and growing
contamination finally causes the base line to start coming down.
The initial base line adjustment at 5 to 6 volts, prevents the maximum base line value from
reaching 9.5V, where the error message “Check base line” appears. Note that this message is
just a warning. The analyzer is not prevented from carrying out further analyses. As long as the
base line remains below 10V, the results are correct.
If a base line is below -9.5V, an adjustment should be done as advised above. If this is not
possible, then most probably there is a fault or a very strong contamination. See chapter 4.1
“
IR
paths cleaning
”. In case of contamination, all cells available should have very low voltages,
even negative ones.
If all available cells are below -9.5V there is most probably a fault that is common for all cells
like no rotating chopper motor or a faulty infrared source. See chapter 2.35
“
The chopper
motor doesn’t rotate
” and 3.3 “
IR source voltage setting
”.
See also Chapter 3.4
“
Analysis takes too long / RF-interference on the IR-cell, or on the
electronic units
”
Pos : 3. 4 / 0025 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 Fluc tuati ng gas flow ELTRA @ 6\ mod_1368452950593_9. doc x @ 40051 @ 2 @ 1
3.2 Fluctuating gas flow
Pos : 3. 5 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 2000_s er vic e/05F aul ts_C S-2000/0510 M odul Fluc tuati ng gas flow @ 9\ mod_1425539177355_9. doc x @ 74105 @ @ 1
Fig. 4: Fluctuating gas flow
The flow rate is electronically controlled; therefore there should be no significant variation from
the set value.
6
7
5
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...