Adjustments
71
Jumper settings:
Fig. 56: Jumper settings
The connector J2 is for selecting different options for the induction generator control. By default,
one jumper is set in the factory at the position “current control” connecting the pins B1-B2 and a
second jumper is set to connect B3-
B4 for “On board power adjustment”.
The connector J3 is for selecting options for the oxygen flow. The factory setting of the jumper is
for “External flow enable”, by connecting the pins A1-A2. In this case, the microcontroller board
can start and stop the flow enabling gas saving mode.
When moving the jumper to B1-
B2 for “On board flow enable”, the flow will start and stop with
the analysis begin and the analysis end, in other words, there is flow only during analysis. This
can be interesting for older analyzers without gas saving function in their software, when they
are used in automations running analyses only once in a while with long and regular idling
period between analyses, for example 15 minutes
At position “Continuous flow” when the jumper connects B3-B4, there is continuous flow as long
as the mains power switch is set to position 2. The continuous flow position is useful in case of
looking to solve flow problems, because the flow is always enabled
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...