Faults
57
CAUTION
Switch off the mains power when handling inside the analyzer.
If a tube fitting is blocked, then the tube may be blocked as well.
Therefore the tube needs to be cleaned or replaced, otherwise the contamination could reach
the IR-cell as well.
2.
The IR-cell is blocked
This fault is caused mainly by a kink in one of the tubes.
A blockage due to contamination may appear in one of the fittings of the IR paths, where
is a minimum diameter in the gas flow, compared to the inner diameter of the tubes.
3.
Leakage
Not only blockages of the flow can cause the pump to run with full power, but also
leakages between the flow sensor (37) and the pump (29) can cause the pump to run at
maximum power.
Occasionally, the cover of the attenuator (30) (see also below) is not properly sealed, or
the gas flow meter (20) has a crack at the fittings for the tube connections. Disconnect
the gas flow gauge and check for cracks.
4.
Blockage in the lower connection of the thick glass tube (22)
If there is not enough quartz wool used underneath the magnesium perchlorate, fine
particles fall down to the glass fitting causing blockage in its hole.
–
Remove the glass tube (22). See operation manual
“
Filling the reagent tubes”.
–
Remove the tube from the lower fitting.
–
Clean the hole by passing a wire or a pipe cleaner.
5.
Cotton wool is dirty
Another possibility could be that the cotton wool inside the glass (30) is very dirty.
CAUTION
Don’t interchange the tubes of the glasses
Fig. 44: Gas pump/ Attenuator
The end of the tube inside the glass must be
free. It shouldn’t be blocked by cotton wool. The
same applies for the fitting at the outlet of the glass.
The tubes of the pump are connected to the short fittings of the glass.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...