Faults
35
Pos : 3. 14 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 D evelopment of dus t when cl osing the fur nac e ELTR @ 6 \ mod_1368620567231_9. doc x @ 40336 @ 2 @ 1
3.7 Development of dust when closing the furnace
Pos : 3. 15 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0540 M odul D evelopment of dus t when cl osing the fur nac e @ 9\ mod_1425539653510_9.doc x @ 74177 @ @ 1
Fig. 26: Gas-flow diagram - furnace closed
This happens when there is no restrictor in the left hand fitting of the valve V3 or when the
internal diameter of the restrictor is too large. This can easily be checked when removing the
connected tube.
The restrictor should have an internal diameter of 1 mm.
See draft
“Gas-flow diagram - furnace closed”
in the chapter
“
Modification Overview
”.
The sample material is causing dust:
This will be caused mainly through tin analysis. Tin causes a light grey dust. The same problem
appears when tin is used as accelerator. Therefore using tin as accelerator should be avoided.
Also the tungsten
– tin mixture shouldn’t be used. Almost all different materials can be analyzed
without the need of tin accelerators.
See Operatio
n Manual chapter “
Applications
”.
Usage of other materials:
Cast-iron causes brown dust during the analysis.
When the dust is yellow, too much tungsten is used as accelerator. If the dust is brown, too
much iron accelerator is used.
See Operation Manual
chapter“
Applications
”.
Pos : 3. 16 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Erratic res ults @ 8\ mod_1396882131370_9.doc x @ 60214 @ 2 @ 1
3.8 Erratic results
Pos : 3. 17 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0545 M odul Erratic res ults @ 9\ mod_1425539699891_9.doc x @ 74189 @ @ 1
There are a lot of different reasons for this
–
Saturated chemicals or clogged dust filter
–
Heterogeneous samples
–
Leakage
–
Flow problems
–
The electronic noise or drift of the infrared signal
–
It can happen quite often that the combustion start is delayed and the
analysis
time is too long
–
Incomplete or bad combustion
–
Too high blank value for the crucible and tungsten. This will be especially
noticeable,
mostly though, in the low carbon channel
–
Widely different weights of sample and accelerator (tungsten)
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...