Faults
21
frequent problem. Tighten the screws of the plug properly, to ensure a proper connection to the
board.
2.
Checking the generator tube
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Ensure that the power switch is at position 3.
Check if the filament of the generator tube glows:
This can be easily checked:
Through the air vent holes of the furnace, the reflected light of the glowing filaments (heater of
the cathode) should be visible.
Fig. 13: Air vent holes of the furnace
3.
The
filament of the generator tube doesn’t glow
Check the voltage of the filament, on the transformer, there should be 10 V AC on the
secondary. If not, then measure 220 V AC on the primary, leads 2-3.
NOTICE
Set the power switch to pos. 1 before removing the tube, although it would be best to
completely turn off the device and unplug the power cord.
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Before touching any components or wiring, make sure the capacitors are discharged
.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...