If the Analyzer vibrates when the mains power switch is turned to position 3, the reason is dust
accumulation on the blades of the oscillator’s blower. The vibrations are caused by dust on the
blades causing eccentrical rotations.
DANGER
Mortal danger from electric shock
Clean it carefully with a vacuum cleaner, if necessary:
Fig. 50: Oscillator blower cleaning
Pos : 3. 62 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Anal ysis c ycle st ops t oo earl y @ 9\ mod_1425626989348_9.doc x @ 75188 @ 2 @ 1
3.31 Analysis cycle stops too early
Pos : 3. 63 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0670 M odul T he anal ysis c ycle st ops t oo earl y @ 9\ mod_1425541597042_9. doc x @ 74501 @ @ 1
The minimum analysis time is set too short in the configuration window, so that the analysis time
expires before the peak starts coming up. Normally the peak starts about 6 to 8 seconds after
the start of combustion. If the minimum analysis time is set for example to 4 seconds, the
analysis may be terminated before the peak starts. If the minimum analysis time is set far over 8
seconds, the peak starts before the minimum time expires. Due to the IR signal being then
higher than the comparator level, the analysis continues until the IR signal comes down to the
comparator level, unless the maximum analysis time expires earlier. If so, it is advisable to
increase the maximum analysis time in order to integrate the whole peak.
Va
c
u
u
m
c
le
a
n
in
g
Va
c
u
u
m
c
le
a
n
in
g
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...