Adjustments
82
13.
Adjust the outputs of other cells (not golden paths) to have between 7 to 8V by adjusting
the gains of their preamplifiers.
14.
Set the jumper of the JP1 from position 2 (Test °C) to Position 3 (Analysis).
15.
Install the IR module into the analyzer.
16.
Allow sufficient time for the base lines to stabilize.
NOTICE
It is necessary to turn slowly in order for the IR sources to stabilize their temperature
with increasing IR source voltage.
–
Adjust the other cells of the module (if any) at 8V baselines or 5V to 6V in case of clean
golden paths.
–
Install the IR module into the analyzer. Allow sufficient time for the base lines to stabilize.
–
Start gas flow and check the base lines again. If any of them is significantly over 8V then
readjust.
Drift too large
–
Check the temperature regulation.
–
Unstable power supply.
–
Defective IR-source.
NOTICE
If it is shown on the display that the infrared cell's output voltage is over 10 volts then
there is a serious defect.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...