Faults
65
Especially in case of resistance furnace operation, the peak of hard-to-burn samples starts
coming up quite late. In such cases, the minimum analysis time should be set to about 60
seconds.
If after prolongation of the minimum analysis time to a reasonable value, a peak still doesn't
appear, possible reasons for the fault are:
–
No combustion. See chapter
“
No combustion
”.
–
There is no gas flow. See chapter
“
No flow
”.
–
There is gas flow, but bypassing the furnace. See chapter
“
Sample burns but no peak
on the screen
“.
–
The IR cell is faulty. See chapter
“
Check base lines
“.
Pos : 3. 64 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 C hopper
motor does n‘t r otate ELTR A @ 6\mod_1368709089073_9.doc x @ 41319 @ 2 @ 1
3.32
Chopper motor doesn‘t rotate
Pos : 3. 65 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0675 M odul T he c hopper mot or does n`t rotate @ 9\ mod_1425541638117_0. doc x @ 74513 @ @ 1
The rotation of the chopper motor needs to be checked in case of problems only, like when
there are no combustion peaks on the screen or the IR signals are minus -10V (take care,
minus) displayed on the
PC screen in the window “inputs”.
In order to check whether the chopper motor is rotating, the easiest way is to remove an IR
path. See chapter “
IR-paths, cleaning and replacing
”. Mostly, the easiest path to remove is
the shortest one. Then, the chopper blade can be seen thru the hole of the chopper housing,
which becomes free (open) after removing the path.
The reason of failure can be a faulty motor but also the controlling electronics of the motor may
have become defective. If so, the motor or the IRC board has to be replaced. Before replacing,
check at least the voltages on the board, mainly the supply voltage of 24VDC. See chapter
“
Infrared cell adjustments
”.
For replacing the chopper see chapter “
Chopper
”.
Pos : 3. 66 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Sample burns but no peak on the s creen @ 7\ mod_1381400077047_9. doc x @ 49828 @ 2 @ 1
3.33 Sample burns but no peak on the screen
Pos : 3. 67 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0680 M odul Sample burns but no peak on the s creen @ 9\ mod_1425541684636_9. doc x @ 74525 @ @ 1
When the analysis cycle is carried out normally and the sample is burned but no peak on the
screen, possible reasons are the following:
-
The IR cell could be faulty. Check the base lines. See chapters “
Check base lines
” and
“
Infrared cell adjustments
”. If the base lines look normal, the IR cell can be regarded as
ruled out. If the base lines are practically at zero volts, the cell is faulty or without power.
(24VDC). See chapter
“
Bad combustion
”.
-
The plunger of bypass valve V4 jams. See
chapter “
Cleaning solenoid valves
”.
This is related to induction operation only:
This may happen in case of induction furnace operation. There is gas flow, but bypassing the
furnace, so that there is gas flow shown on the flow meter and by the flow sensor on the screen,
but no gas flow thru the induction furnace. In this case the bypass valve V4 jams, causing the
oxygen f
low to bypass the furnace like in case of open furnace. See chapter “
Gas flow
system
”.
The furnace has though oxygen pressure but without having oxygen flow. Due to oxygen being
present, the metal sample and accelerator(s) burn, because there is availability of oxygen in the
furnace, coming thru the chamber valve V3 and/or thru the lance valve V2. In case of metal
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...