5-4
Installation, Setup, and Operation
CESAR 1312 (E)
Dressler
®
HF-Technik GmbH
Grounding
Connect the ground terminal at the rear panel of the CESAR generator with
the nearest PE ground (usually the 19” rack). Use a proper grounding cable
that meets your national standards and has a cross-sectional area of at least
4 mm
2
. Proper operation requires that you use a central, common ground for
the generator, matching network, and load.
Connecting RF Output Power
The following illustration shows the output power connector, and
describes each pin.
Figure 5-1. Output (RF Out) connector
To Connect RF Output:
1. Connect one end of a shielded coaxial cable of 50
Ω
impedance to the
RF
Out
connector.
Do not attempt to turn on power until the CESAR generator is grounded.
Table 5-1. Output connector pin descriptions
Pin
Description
Center
RF output
Outer
Ground