CESAR 1312 (E)
Installation, Setup, and Operation
5-1
Chapter
CESAR™ 1312 Generator
Chapter
5
Installation, Setup, and
Operation
PREPARING TO INSTALL THE CESAR
GENERATOR
The following sections provide information that you need to understand before
installing the CESAR generator.
Spacing Requirements
The CESAR generator is designed for 19" racks but may be used alternatively
as a desktop model. Regardless, you must ensure proper air flow (see
“Cooling Specifications” on page 3-6
• 6 cm (2.4") required on the left and right sides for airflow
• 10.16 cm (4
″
) required at rear for cable connections
Additional space for connectors and cabling can be necessary. Also, allow
adequate space for anyone who may need to access the front and rear
panels. For a dimensional drawing, see