CESAR 1312 (E)
Communication Interfaces
4-17
CESAR™ 1312 Generator
7
20
DC BIAS INPUT
Analog
input
This input signal is used for
DC bias regulation, where the
set point is given by pin 5
(scaling must be the same).
The signal closes the control
loop around external
components in the RF path.
The default range is 0 V to
10 V, but you can change this
range (see
Default setting: 0 V to 10 V =
0 V to 4000 V.
Typically, matching networks
provide a DC bias voltage
monitor signal. When this
scaled representation of the
DC bias voltage is used for
DC bias regulation (see pin
6), the scaling must be the
same as for the set point (see
pin 5). Dressler Variomatch
matching networks provide a
DC bias voltage monitor
signal for regulation through
the
Matching Interface
.
For a wiring diagram, see
.
Table 4-2. 25-pin User Port pin descriptions (Continued)
Pin
Number
Return
Pin
Name
Signal
Type
Description
Содержание CESAR 1312
Страница 1: ...U s e r M a n u a l s e r v i c e d r e s s l e r c o m CESAR Generator Model 1312 ...
Страница 2: ......
Страница 3: ...User Manual CESAR Generator Model 1312 CESAR 1312 E ...
Страница 6: ...iv CESAR 1312 E Dressler HF Technik GmbH ...
Страница 14: ...xii List of Tables CESAR 1312 E Dressler HF Technik GmbH ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines CESAR 1312 E Dressler HF Technik GmbH ...
Страница 34: ...3 8 Specifications CESAR 1312 E Dressler HF Technik GmbH ...
Страница 170: ...5 50 Installation Setup and Operation CESAR 1312 E Dressler HF Technik GmbH ...
Страница 188: ...6 18 Troubleshooting and Customer Support CESAR 1312 E Dressler HF Technik GmbH ...