4-34
Communication Interfaces
CESAR 1312 (E)
Dressler
®
HF-Technik GmbH
10
8
RF POWER
ON
Digital
input
This signal allows you to enable or
disable RF power.
For a wiring diagram, see
11
8
DC BIAS SET
POINT
Analog
input
This signal provides a linearly
scaled setting to the DC bias
voltage. The default range is 0 V to
10 V, but you can change this
range (see
)
0 V to 10 V = 0 V to maximum DC
bias (4000 V default maximum
value)
For a wiring diagram, see
12
8
RF POWER
SET POINT
Analog
input
This signal provides a linearly
scaled setting to the RF forward
power. The default range is 0 V to
10 V, but you can change this
range (see
)
For example, 5 V = 50% of the
nominal generator power; 10 V =
100%.
For a wiring diagram, see
13
8
TEST
VOLTAGE
FOR
FORWARD
POWER
Analog
output
This signal provides a linearly
scaled readback of RF forward
power as measured at the
generator output. The default range
is 0 V to 10 V, but you can change
this range (see
0 V to 10 V = 0 W to “XXX kW”
For a wiring diagram, see
Table 4-4. 15-pin User Port pin descriptions (Continued)
Pin
Number
Return
Pin
Name
Signal
Type
Description
Содержание CESAR 1312
Страница 1: ...U s e r M a n u a l s e r v i c e d r e s s l e r c o m CESAR Generator Model 1312 ...
Страница 2: ......
Страница 3: ...User Manual CESAR Generator Model 1312 CESAR 1312 E ...
Страница 6: ...iv CESAR 1312 E Dressler HF Technik GmbH ...
Страница 14: ...xii List of Tables CESAR 1312 E Dressler HF Technik GmbH ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines CESAR 1312 E Dressler HF Technik GmbH ...
Страница 34: ...3 8 Specifications CESAR 1312 E Dressler HF Technik GmbH ...
Страница 170: ...5 50 Installation Setup and Operation CESAR 1312 E Dressler HF Technik GmbH ...
Страница 188: ...6 18 Troubleshooting and Customer Support CESAR 1312 E Dressler HF Technik GmbH ...