6-8
Troubleshooting and Customer Support
CESAR 1312 (E)
Dressler
®
HF-Technik GmbH
Error Descriptions and Resolutions.
lists the error messages that may occur on the CESAR generator,
describes each error message, and explains how to resolve the error. If the
information in this table does not resolve the problem, contact Customer
Support (see
“Customer Support” on page 6-14
).
Table 6-1. Error description and resolution
Error Message
Description and Resolution
E01 Interlock Loop
Open
Open interlock loop.
Causes: Switch opened, interlock loop not connected.
Solution: Close interlock loop by making the proper
connections to satisfy the interlock.
• If you have a 25-pin
User Port
, see pins 10 and 23 in
.
• If you have a 15-pin
User Port
and 9-pin
Interlock
connector, ensure external contact closure between
pins 1 and 2 of the
Interlock
connector (see
Note:
To determine which communication interfaces
you have, see
E10 SMPS Temperature
Too High
Switch mode power supply temperature is too high.
Cause: Ambient temperature too high / air flow too low.
Solution: Let generator cool down, improve cooling
Ensure proper airflow:
1. Check to see that the air flow is unobstructed and
below the specified input temperature (see
).
2. Listen to ensure that the fan is operating properly.
Содержание CESAR 1312
Страница 1: ...U s e r M a n u a l s e r v i c e d r e s s l e r c o m CESAR Generator Model 1312 ...
Страница 2: ......
Страница 3: ...User Manual CESAR Generator Model 1312 CESAR 1312 E ...
Страница 6: ...iv CESAR 1312 E Dressler HF Technik GmbH ...
Страница 14: ...xii List of Tables CESAR 1312 E Dressler HF Technik GmbH ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines CESAR 1312 E Dressler HF Technik GmbH ...
Страница 34: ...3 8 Specifications CESAR 1312 E Dressler HF Technik GmbH ...
Страница 170: ...5 50 Installation Setup and Operation CESAR 1312 E Dressler HF Technik GmbH ...
Страница 188: ...6 18 Troubleshooting and Customer Support CESAR 1312 E Dressler HF Technik GmbH ...