CESAR 1312 (E)
Communication Interfaces
4-65
CESAR™ 1312 Generator
9
maximum
external
feedback
Specifies the external feedback value
that corresponds to the maximum voltage
at the
User Port
.
Send three data bytes, least significant
byte first.
• Bytes 1 and 2 = Maximum external
feedback value in the range of 100 to
4000 (100 V to 4000 V).
• Byte 3 is only for compatibility and is
ignored.
3
1
(CSR only)
11
select
active
target
Specifies which target is active.
Send one data byte.
• Accepts a value of 0 to 4.
Note:
If 0 is selected, the no target
life timer will count.
For more information, see
.
Read back with command
156
. (See also
command
12
.)
1
1
(CSR only)
12
AE Bus and
Ethernet
version
set target
life
(see next
row for
Profibus
version of
command)
Sets the target life, in kilowatt hours, of
the specified target (see command
11
).
Send five data bytes, least significant
byte first.
• Byte 1 selects the target number (1
to 4).
• Bytes 2 through 5 specify the target
life in hundredths of kWh (for
example, 100 = 1 kWh).
For more information, see
.
Read back with command
157
.
5
1
(CSR only)
Table 4-16. Host port commands (Continued)
Command
Description
Number of
Host Data
Bytes
Number of
Response
Data Bytes
Содержание CESAR 1312
Страница 1: ...U s e r M a n u a l s e r v i c e d r e s s l e r c o m CESAR Generator Model 1312 ...
Страница 2: ......
Страница 3: ...User Manual CESAR Generator Model 1312 CESAR 1312 E ...
Страница 6: ...iv CESAR 1312 E Dressler HF Technik GmbH ...
Страница 14: ...xii List of Tables CESAR 1312 E Dressler HF Technik GmbH ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines CESAR 1312 E Dressler HF Technik GmbH ...
Страница 34: ...3 8 Specifications CESAR 1312 E Dressler HF Technik GmbH ...
Страница 170: ...5 50 Installation Setup and Operation CESAR 1312 E Dressler HF Technik GmbH ...
Страница 188: ...6 18 Troubleshooting and Customer Support CESAR 1312 E Dressler HF Technik GmbH ...