Non-invasive blood pressure (NiBP)
Start monitoring
The main steps can be undertaken at the parameter box.
1
Plug the hose connectors into the parameter box.
The rubber rings on the connectors can be moistened
slightly so that they can be inserted more easily.
●
Apply the appropriate cuff to the patient and connect
it to the hose. Ensure that all connections are tight.
Even slight leaks make measurement impossible.
Manual measurement
2
Press
‚
.
The cuff is inflated and the cuff pressure indicated
constantly in the bar graph within the module.
3
The systolic, diastolic and
4
mean pressure are displayed when the measurement
is complete.
Incorrect measurements are repeated 5 seconds after
releasing the air. A third and last attempt is made after
30 seconds if the second measurement also proves
incorrect.
Automatic measurement
5
Time bar indicating the relative time elapsed between
two automatic measurements. The actual cuff
pressure is indicated here during the measurement.
6
Time interval for automatic measurement.
7
Press
¡
.
The LED in the key lights up and automatic measure-
ment starts. NiBP measurements are performed by
the parameter box at the set time intervals.
To interrupt measurement
2
Press
‚
.
to interrupt measurement. The air is immediately
expelled from the cuff.