Message, cause and remedy
Messages on the system screen
(In alphabetical order according to priority)
Messages
Cause
Remedy
Advisory messages
AGAS MIX !!
Gas measurement has detected a mixture of two
anaesthetic agents. The leading agent is
recognised and quantified. The accuracy of this
process is limited.
If necessary, check the
anaesthetic agent concentration
in the fresh gas.
AGAS INOP !
Anaesthetic gas measurement faulty.
Call DrägerService
BP CUFF ? !
Cuff hose disconnected?
Leak in hose or cuff?
Check connections.
Replace cuff or hose.
CAN COM INOP !
Communication via CAN interface faulty.
Check connections of data lead.
Call DrägerService
CHECK IBP 1 !
CHECK IBP 2 !
No pulse detected by pressure measurement.
Check transducer, replace if
necessary.
CHECK PRESSURE !
No breathing phase detected by pressure
measurement.
Check pressure measurement.
CO2
G
OFF !
CO
2
alarm inactive.
CO2/AGA INOP !
CO
2
/ anaesthetic gas measurement faulty.
Call DrägerService
CO2 INOP !
CO
2
gas measurement faulty.
Call DrägerService
CO2 LINE ? !
Sample line blocked.
Check sample line, filter in
T-piece and water separator in
water trap; replace if necessary.
Check sample line is not kinked.
Use the Dräger sample line.
CO2 WARMUP !
The CO
2
sensor has not yet attained the required
temperature with complete accuracy.
Wait (max. 6 minutes).
COOLING 8060 ? !
Temperature inside machine too high.
Clean filter at rear.
Call DrägerService
ECG INOP !
or
ECG NOISE !
The ECG signal is encountering interference due to
electrosurgery.
The heart rate and S-T segment analysis will be
frozen for 15 seconds and deactivated if the
interference persists.