78
System screen functions
Alarm concept
11:59
00:00
06.12.98
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
10.2
29
MV l/min
99
SpO
2
%
Freq 12 V
T
0.60
∆
O
2 --
Fi O
2
Vol %
T2
37.5
°
C
T1
36.8
°
C
33
etCO
2
mmHg
1/min
33 10.2 19
et CO2 MV Peak
12 0.81 1
Freq. VT PEEP
72
Peak mbar PEEP
32 3
Plat 18 Mean 6
HR/Puls
72
50 120
Pulsdef.
4
5
NiBP sys
122
80 160
SpO
2
92
91 --
et CO
2
33
30 50
in CO
2
--
5
AMV
6.2
3.0 12.0
Fi O
2
29
20 --
Fi SEV
0.8
-- 3.4
PAW
9
8 40
iBP 1 sys
125
80 167
iBP 2 mit
31
0 8
ST
-0.05
±
0.10
T1
36.8
-- --
T2
37.5
-- --
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
1. ECG I
pulse def. 4 ST + 0.05
auto set
pat. alarm.
timer
start
alarm
info
auto set
vent. al.
Select the required limits and confirm !
HR/Pulse (ECG)
limits
79 99 --
/
--
HR/Pulse SpO2 Art (1)
1/min % mmHg
16:47
00:00
01.12.98
P. Alarme
anpassen
V. Alarme
anpassen
Stopuhr
Start
Para-
meter
Konfig.
Alarm
Info
Bildsch.
Aufbau
Grenzen
IPPV Alarmgrenzen
Standardbildschirm 1 - Erwachsene
Alarm:
Achtung:
Hinweis:
1
2
3
4
14129278
05829278
Alarm concept
Priority of alarms
The alarms are arranged in order of priority and assigned
to specific tones or tone sequences.
The information appropriate to the situation is clearly
presented in the separate fields for:
Warning
marked »
!!!
«
Text in a red field
Caution
marked
»
!!
«
Text in a yellow field
Advisory
marked
»
!
«
Text in a white field
Warnings are accompanied by a continuous tone, caution
messages by a tone at 30 second intervals and advisory
messages by a single tone. Low-priority technical
messages are displayed without audible alarm tone.
Tones conforming to the EURO standard or Dräger
standard are available as alternatives.
For warnings, the tones conforming to the Euro standard
are separated by pauses of about 15 sec.
4
Whenever a warning or caution message occurs,
the corresponding LED lights up above the
G
key:
Warning
red,
flashing
Caution
yellow,
flashing
Advisory
yellow, constant
A summary of all alarm limit settings appears on the
screen, and the parameter which has been exceeded is
highlighted.
Example:
The SpO
2
concentration has fallen below the lower
alarm limit for SpO
2
, which is set to 91%.
The measured value is 90%.
– In the alarm field, the symbol for the lower alarm limit
<
and the parameter name »
SpO
2
« are displayed.
The field changes colour from blue to red.
– The red LED flashes.
– The alarm tone is sounded.
– The alarm limits menu is automatically opened. The
upper or lower alarm limit that has been exceeded or
not attained is highlighted.
This value can be altered by turning the rotary control
and confirmed by pressing the control.
The alarm limit can be confirmed directly if it is to
remain unchanged.
The limit value menu is always closed whenever a
value is confirmed. If the alarm condition is no longer
present, the limit value menu closes automatically.