![Dräger Cicero EM Скачать руководство пользователя страница 128](http://html1.mh-extra.com/html/drager/cicero-em/cicero-em_instructions-for-use-manual_2526452128.webp)
Connect fresh gas hose
1
to breathing system from below with plug connector.
Connect pressure measuring line with filter
2
Plug pressure measuring line into coupling - until it
engages.
Fill soda lime container
The container can hold approx. 1.5 l soda lime
»
Drägersorb 800 plus
« (see separate Instructions for
Use). This is sufficient to bind approx. 150 l CO
2
or
for a period of at least 6 hours.
●
Do not fill the soda lime until just before using!
Soda lime that has been exposed for a long time to the
ambient air and/or radiant heat dries out and therefore
loses some of its capacity to bind CO
2
. The efficiency of
the indicator substance (change of colour) is also
reduced.
The soda lime should be replaced daily, at the latest
when two-thirds of the charge has changed colour.
The indicator colour
changes back as it dries
although
the absorption capacity is exhausted.
Absorber function
●
Place the chain on the bottom of the container.
●
Insert mesh -
must not be forgotten! -
Absolutely essential for absorption!
●
Uniformly fill the container with soda lime all round,
until the
– MAX –
mark is reached (approx. 3 cm below top edge).
●
Gently shake the container or knock against the table
to settle the contents.
●
Remove any dust and granulated material on the edge
of the container.
Fit the soda lime container
3
into the breathing system from below and turn
clockwise as far as possible.
128
D
Reassembling machine
Reassembling the parts
1
– MAX –
3
APL
1
2
M
A
N
M
A
N
2
18329278
18429278
06429278