SpO
2
/Pleth settings
(see also page 97):
measuring function
Switches SpO
2
measurement on and off.
pulse
Switches pulse rate display on and off.
C-Lock
Switches ECG synchronization on and
(ECG synchr.)
off.
mode
Determines the speed of measurement.
Fast measurements are more susceptible
to antefoets.
CO
2
/O
2
settings
sample rate
Selects the flow rate of the sample gas
through the sample gas line at the Y-
piece. Dräger recommends 200 mL/min
with connected sample gas return line.
CO
2
amplitude
The maximum amplitude of the CO
2
real-
time curve can be set here.
CO
2
unit
The CO
2
measuring unit can be selected
between mmHg and kPa.
insp. CO
2
Switches the numerical display of the
inspiratory CO
2
concentration on or off.
Inspiratory CO
2
is monitored in IPPV
mode with an adjustable alarm limit of
0 to 10 mmHg.
sidestream O
2
measurement
»on«
:
O
2
is measured by sampling the
sidestream.
»off«
:
O
2
is measured by measuring
O
2
at the inspiratory port
(requires inspiratory O
2
-sensor).
pEEG settings
This menu selects the pEEG Parameters displayed in the
pEEG module (see page 58).
52
System screen functions
Configuring the system screen in »Standby«
Standby / Configuration
11:59
00:00
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
Select a menu position. Please confirm !
Change the settings. Please confirm again !
alarms inactive !
ECG
NiBP
iBP locations
iBP channels
SpO
2
/ Pleth
CO
2
/O
2
pEEG
CO
2
/O
2
sample rate
ml/min
60 200
amplitude
mmHg CO
2
0-75 0-110
units CO
2
mmHg kPa
Vol.%
insp. CO
2
on off
sidestream
on
off
O
2
-measurem.
06-12-98
Standby / Configuration
11:59
00:00
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
Select a menu position. Please confirm !
Change the settings. Please confirm again !
alarms inactive !
ECG
NiBP
iBP locations
iBP channels
SpO
2
/ Pleth
CO
2
/O
2
pEEG
pEEG
SEF / SMF / EMG
SEF /
δ
.Ratio / EMG
L.SEF / R.SEF / EMG
SMF /
δ
.Ratio /EMG
SEF /
δ
.Ratio /
δ
/
θ
/
α
/
β
L.SEF / R.SEF / SMF
Amplitude
06-12-98
Standby / Configuration
11:59
00:00
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
Select a menu position. Please confirm !
Change the settings. Please confirm again !
alarms inactive !
ECG
NiBP
iBP locations
iBP channels
SpO
2
/ Pleth
CO
2
/O
2
pEEG
function
on off
pulse
on off
C-Lock
(ECG
synchr.)
on off
mode
slow normal
fast
SpO
2
/Pleth
06-12-98
08629278
08729278
08829278