Changing patients
Changing parts:
Refer to page 117 and 125 for a schematic overview and
description of methods for cleaning and disinfection.
Maintenance intervals: page 137.
●
Machine in Standby mode -
●
Vapor: thumbwheel set to »
O
« = Off.
After treating an infectious patient
the entire machine must be cleaned, disinfected and
sterilized.
Note hygiene regulations -
risk of infection!
After treating a non-infectious patient
the following parts must be replaced before continuing
with the next patient:
●
Tube or mask
●
Y-piece
●
Both breathing hoses
●
Temperature sensor and cable if applicable
1
T-piece of the measured gas hose and filter, if used
(disposable articles, household refuse) -
2
Draw the water trap container down and off to be
emptied -
The following must also be changed when working
without microbial filters:
3
O
2
sensor with connecting lead, if used.
4
Flow sensor.
Disconnect lead from flow sensor - it remains on the
machine.
●
Breathing bag with hose
●
Breathing system:
Disconnect fresh gas hose - it remains on the machine.
●
Pressure measuring hose and filter - must be
replaced. Hose connector remains on the machine.
●
Soda lime container - turn anticlockwise and pull
down - must be replaced.
●
Release the hose of the anaesthetic gas scavenging
system (lock in front of the connector - No. 5 on
page 118), and also disconnect from the wall socket
of the central scavenging system.
43
FULL
Anaesthesia ventilation
Changing patients
1
2
4
3
06729278
06829278
06929278