Installing the anaesthetic gas scavenging
system:
●
Hook the collecting system to the mount on the back
of the Cicero:
Place the slots of the receiving system over the
holders and slide the receiving system down.
See also
"What's what"
on page 146, No. 14.
Connecting the transfer hoses
1
Pass transfer hose from behind through the hole in
the Cicero EM and
2
connect the hose to the scavenging connector of the
breathing system (only on first installation - the hose
then remains on the connector).
3
Connect the transfer hose to the connector on the
receiving system.
●
Make sure the connection is secure!
4
Make sure that the connector for the second transfer
hose is sealed with a screw plug.
Do not plug the openings of the receiving system, other-
wise the breathing system could be drained.
5
Connect the scavenging hose to the output connector
of the scavenging system.
6
Plug the connector of the scavenging hose into the
wall socket. The visual indicator shows »
green
«.
18
Preparing for use
AGS-system
Bronchial aspirator
D
1
5
3
D
2
1
Detailed view
from below after
removing soda
lime container.
4
6
02129278
02229278
02329278