Dismantling parts
Measured gas:
1
T-piece of measured gas hose (if used): remove filter.
Inspiratory O
2
sensor
(if used)
:
●
Must not be disinfected in liquid or autoclaved!
●
Wipe any dirt off housing or lead with damp
disposable cloth.
●
Unscrew O
2
sensor and remove O
2
cell. Do not
touch wire mesh of O
2
cell, otherwise it loses its
water-repellent property and measurement becomes
inaccurate.
Dirt on wire mesh of sensor cell is simply wiped off
with a disposable cloth moderately soaked in distilled
water.
Temperature sensor:
(breathing gas; if used)
●
Draw temperature sensor out of Y-piece, remove
Y-piece from ventilation hoses.
Prise lead out of hose clips.
●
Pull connector out of socket.
●
Wipe dirt off with damp disposable cloth.
Flow sensor
Contamination by infectious patients: dispose of as
directed!
If there is no risk of contamination, continue using the
flow sensor until the message »
FLOW INOP !
« appears at
the breathing minute volume. Then dispose of the flow
sensor as ordinary domestic waste.
It can be incinerated at temperatures above 800
°
C with
low pollutant emissions.