Check manual ventilation function
Machine in Manual/Spontaneous mode:
1
Press
M
-
2
Pressure limiting valve (APL) set to MAN -
●
Connect lung simulator, test thorax or breathing bag
to Y-piece -
3
Set fresh gas flow -
2
Set maximum ventilation pressure between 5 and
70 mbar on APL valve - turn valve head for this
purpose.
●
Compress breathing bag -
●
Compare pressure indication on system screen with
setting on pressure limiting valve.
If excess pressure has to be relieved quickly:
2
press vertically down on the lever of the pressure
limiting valve.
Check automatic ventilation function
●
Connect lung simulator, test thorax or breathing bag
to Y-piece -
3
Set fresh gas flow -
4
Press
I
-
5
Press rotary control -
6
Piston movement is indicated on the bar graph.
The machine starts with the ventilation parameters
set upon delivery (or programmed to customer's
requirements by DrägerService).
– Lung simulator inflates regularly.
– Pressure profile is displayed on the system screen.
– Volume measurement yields plausible values.