Non-invasive blood pressure (NiBP)
Input.........................................: IEC type CF, floating, Colour-coded purple
Measuring method ....................: Oscillometric
Measuring range.......................: Adult mode:
20 to 290 mmHg
(3.0 to 38.6 kPa)
Neonatal mode:
5 to 150 mmHg
(0.6 to 20 kPa)
Resolution: ...............................: 1 mmHg (0.1 kPa)
Accuracy ..................................: 1 mmHG (0.1 kPa)
Unit of measure ........................: mmHg or kPa
Inflation of cuff ..........................: The cuff is inflated to the preset starting pressure for the first measurement; after
reading off the systolic pressure, the cuff is inflated to 20 mmHg above the last
measured systolic value for the next measurement. If the inflation pressure is too low
for correct measurement of the systolic pressure, air is automatically released from
the cuff and the measurement is repeated 5 seconds later with a higher cuff
pressure. If it is still too low, the air is released again and the measurement repeated
30 seconds later with an elevated cuff pressure.
Release air from the cuff ...........: – Upon completion of the measurement.
– When measurement is interrupted by pressing »
Stop
« button.
– When the maximum measuring time is exceeded.
– When the maximum cuff pressure is reached.
– In the presence of undue artefacts.
– In the event of a power failure / when switching off the monitor.
– When NiBP function deactivated.
Max. cuff pressure ....................: 300 mmHg
(40 kPa)
for adults
150 mmHg
(20 kPa)
for neonates
Max. measuring time .................: 90 seconds
for adults
60 seconds
for neonates
Intervals for automatic
measurements ..........................: Every 2, 3, 5, 10, 15 minutes. Manually triggered interim measurement
does not affect the time between automatic measurements.
Venous congestion ...................: Adjustable cuff pressure
Invasive blood pressure (iBP)
Input.........................................: Floating, IEC type CF; colour-coded grey.
Input sensitivity .........................: 42.5 and 50
µ
V/V/10 mmHg
Input impedance .......................: 90 k
Ω
(can be changed to 1 M
Ω
)
Bridge voltage ..........................: 5 V DC
Bridge impedance ....................: 200
Ω
to 10 k
Ω
Zero alignment..........................: Semi-automatic; not possible while pressure continues to fluctuate
(with site label »
ART
« or »
AORTA
«);
range
±
70 mmHg. Inaccuracy < 0.5 mmHg.
Bandwidth ................................: 0 to 7 Hz (– 3 dB)
Technical data