ECG / Heart rate
ECG display
●
Insert the electrode cable in the ECG input of the
parameter box, and attach the electrodes to the
patient (for instructions, see »
Descriptions
« from
page 163 onwards).
1
The ECG of the preprogrammed lead appears on the
system screen as specified in the configuration menu.
2
The heart rate is displayed.
Adjusting settings
●
Press the softkey »
Parameter
«. The menu
for setting the parameters is displayed.
●
Move the cursor frame to »
ECG
« with the
rotary control and press to confirm.
measuring
function
on / off
Activates and deactivates the ECG
measuring function.
number
of leads
3 / 5
Selects the type of cable used. If you are
using a 3-lead ECG cable, leads 1, 2 and 3
are available. If only 3 electrodes of a 4 or
5-lead cable are connected, the system
displays the following error message:
»
Elektrode
«.
lead
I
/
II
/
I I I
/ AVR / AVL / AVF / V
Selects the lead separately for the 1st, 2nd
and 3rd ECG. With a 3-core cable, the
choice is restricted to lead 1, 2 and 3.
amplitude
mV / 4 / 2 / 1 / 0.5 / 0.25
Sets the ECG amplitude in mV on the
screen in relation to the reference bar to the
left of the ECG graph. Changes to settings
are immediately visible on the curve.
filter
on / off
»
on
« yields an extremely stable ECG
without interference.
»
off
« yields an ECG with higher band
width, but less stable.
pacemaker on / off
Activates and deactivates pacemaker
pulse detection. Pacemaker pulses are
enlarged disproportionately when this
function is activated (not taken into account
when calculating the heart rate).
Parameter box
ECG / Heart rate
89
g
d
AUTO
Start
Stop
Punct.
Act.
→
0
←
→
0
←
11:59
00:00
06-12-98
Man./Spont. alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
II
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
33
6.0
freq 10 V
T
0.60
36.8 37.2
pulse def. 4 ST + 0.05
65
pulse 65 (100)
Peak PEEP
32 3
plat 26 mean 21
pulse 65
99
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat. alarm.
timer
start
para-
meter
config.
alarm
info
screen
config.
limits
auto set
vent. al.
etCO
2
mmHg
MV L/min
T
1
T
2
°
C
HR/Pulse ECG
1/min
IBP1 ART mmHg
122
/
89
SpO
2
%
1
2
11:59
00:00
Select the required parameter and confirm !
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
2. ECG
II
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
33
6.0
freq 10 V
T
0.60
36.8 37.2
pulse def. 4 ST + 0.05
65
pulse 65 (100)
Peak PEEP
32 3
plat 26 mean 21
pulse 65
99
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat. alarm.
timer
start
config.
alarm
info
screen
config.
limits
auto set
vent. al.
ECG
NiBP
iBP 1
iBP 2
SpO
2
/ Pleth
CO
2
/ O
2
pEEG
para-
meter
etCO
2
mmHg
MV L/min
T
1
T
2
°
C
HR/Pulse ECG
1/min
IBP1 ART mmHg
122
/
89
SpO
2
%
06-12-98
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
2. ECG
II
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
33
6.0
freq 10 V
T
0.60
36.8 37.2
pulse def. 4 ST + 0.05
65
pulse 65 (100)
Peak PEEP
32 3
plat 26 mean 21
pulse 65
99
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat. alarm.
timer
start
alarm
info
auto set
vent. al.
para-
meter
Select a menu position. Please confirm !
Change the settings. Please confirm again !
ECG
function
on
off
number
3
5
of leads
1.ECG 2.ECG 3.ECG
lead I II III
amplitude
1
filter
on
off
pacemaker
on
off
pulse def.
on
off
indication
ST-segment
analysis
on
off
etCO
2
mmHg
MV L/min
T
1
T
2
°
C
HR/Pulse ECG
1/min
IBP1 ART mmHg
122
/
89
SpO
2
%
06-12-98
14929278
15029278
15129278
15229278