Anaesthesia ventilation
Select Manual / Spontaneous mode
1
Press key »
MAN SPONT
« on ventilator for at least
one second (or press the key and confirm with the
rotary control. Setting by Dräger Service as from
software version 7.2).
2
Display in dialogue field:
MAN/SPONT
»Standard screen 1«
appears with the alarm limits for
MAN/SPONT mode.
The
P
and
p
keys on the ventilator are
disabled.
If the manual ventilation bag is squeezed without pressing a key
while the machine is in »
Standby
« mode, the machine
automatically activates MAN/SPONT mode.
Select the carrier gas for the anaesthetic agent:
3
Set the switch to »
N
2
O
« or »
Air
«.
Spontaneous breathing
4
Set pressure limiting valve APL to
»SPONT«
. It is
now open, regardless of the set pressure.
5
Set fresh gas –
detailed information on setting the fresh gas flow
can be found in the Annex on page 168.
Manual ventilation
with breathing bag
The airway pressure is limited via the pressure limiting
valve APL.
4
Set pressure limiting valve APL to
»MAN«
and set
required ventilation pressure:
turn valve head.
To fill system:
6
Press
»O
2
+
«
.
5
Set fresh gas with O
2
, N
2
O or AIR delivery valve –
●
start manual ventilation.
32
Anaesthesia ventilation
Manual / Spontaneous
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
2. ECG
II
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
33
6.0
freq 10 V
T
0.60
36.8 37.2
pulse def. 4 ST + 0.05
65
pulse 65 (100)
Peak PEEP
32 3
plat 26 mean 21
pulse 65
99
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat. alarm.
timer
start
para-
meter
config.
alarm
info
screen
config.
limits
auto set
vent. al.
etCO
2
mmHg
MV L/min
T
1
T
2
°
C
HR/Pulse (ECG)
1/min
IBP1 ART mmHg
122
/
89
SpO
2
%
06-12-98
T2
37.5
T1
36.8
TEST
Lecktest ?
O
2
AIR
MAN
SPONT
IPPV
Pmax
PPCV
SIMV
PCV
VT
fIPPV
TI:TE
fIMV
TIP:TI
Flow
PEEP
1
2
4
5
6
3