pulse deficit indication on / off
Activates and deactivates the pulse deficit
indication derived from the heart rate and
SpO
2
pulse.
ST-segment analysis on / off
Activates and deactivates ST-segment
analysis. Alarms only active in adult mode.
Monitoring of limit values is deactivated
when »
off
«.
The analysis is based on the derivation from
the 1st ECG.
ST-segment analysis may be impaired
by simultaneous high-frequency electro-
surgery!
Electrosurgery during an ECG
ECG measurements are susceptible to interference
from electrosurgery (high frequency surgery). To avoid
unnecessary false alarms, whilst still guaranteeing safe
monitoring, the system responds as follows to detected
high frequency interference:
– the ECG curve is always displayed,
– an »
ECG noise
« error message is displayed,
– the heart rate display is frozen and
– ECG alarms are suppressed.
After the high frequency interference has ceased, the
complete monitoring function is again active. If high
frequency interference lasts longer than 15 seconds
without interruption, the heart rate and S-T segment
values are replaced by dashes (» – – «).
Parameter box
ECG / Heart rate
90
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
2. ECG
II
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
33
6.0
freq 10 V
T
0.60
36.8 37.2
pulse def. 4 ST + 0.05
65
pulse 65 (100)
Peak PEEP
32 3
plat 26 mean 21
pulse 65
99
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat. alarm.
timer
start
alarm
info
auto set
vent. al.
para-
meter
Select a menu position. Please confirm !
Change the settings. Please confirm again !
ECG
function
on
off
number
3
5
of leads
1.ECG 2.ECG 3.ECG
lead I II III
amplitude
1
filter
on
off
pacemaker
on
off
pulse def.
on
off
indication
ST-segment
analysis
on
off
etCO
2
mmHg
MV L/min
T
1
T
2
°
C
HR/Pulse ECG
1/min
IBP1 ART mmHg
122
/
89
SpO
2
%