System screen functions
Configuring the system screen in »Standby«
Select the required ventilation mode (adult or neonate). Please confirm !
Standby / Configuration
11:59
00:00
alarms inactive !
Default values
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
06-12-98
Standard alarm limits configuration menu. Please confirm !
Standby / Configuration
11:59
00:00
alarms inactive !
Settings
mode
adult neo.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
06-12-98
standard adult
standard neonates
Man/Spont-alarmlimits
et CO
2
on off
in CO
2
on off
O
2
high on off
MV on off
Standby / Configuration
alarms inactive !
11:59
00:00
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
160
80
100
50
24
8
24
8
20
--
8
0
16
0
160
80
iBP
Aorta
sys
dia
A.Pulm
sys
mean
dia
CVP
mean
ICP
mean
mmHg
sys
mean
dia
--
120
50
160
80
--
--
--
--
--
--
92
100
50
3,4
--
1,5
2,3
--
3,4
--
--
12,0
--
Anesth. Agent
Fi Hal
Fi Iso
Fi Enfl.
Fi Sevo
Fi Des
12,0
3,0
40
8
50
--
--
20
Pressure /
Ventilation
MV
PAW
etCO
2
FiO
2
Standard - Adult
ECG / SpO
2
/
NiBP / Temp
HR
Pulse
SpO
2
NIBP/ART
sys
dia
T
1
T
2
ST
120
50
Standard alarm limits configuration menu. Please confirm !
100
50
24
8
06-12-98
07829278
07929278
08029278
Mode
The Cicero EM has separate operating modes for adults
(adult) and neonates (neon). These modes differ in the
following way:
– Default alarm limits (see below)
– Volumeter measuring range (see page 58)
– NiBP functions (see page 91)
– ECG functions (see page 89).
Every time the Cicero EM is switched on, it automatically
operates in the mode selected in this menu.
Alarm limits
The standard alarm limits are defined
– separately for adults and neonates,
and, in order to activate/deactivate the MAN/SPONT
alarm limits, for
– et CO
2
,
– in CO
2
,
– O
2
high and
– AMV.
The default alarm limits are automatically activated after
either:
– switching on the Cicero EM,
– leaving
»Standby«
mode,
– selecting
»default«
under
»Alarms«
,
– switching over between
»Neon«
and
»adults«
mode
The limit values are allocated in groups to the following
measurement parameters:
–
iBP
–
ECG / SpO
2
/ NiBP / Temp
–
Anaesth. Agent
–
Pressure / Ventilation
Both the upper and lower limit is displayed at the limit
symbol:
22
Upper limit
11
Lower limit
Two dashes (– –) instead of a number indicate that this
alarm limit is deactivated and not monitored. This is set
by turning the rotary control respectively above or below
the maximum or minimum settable value and pressing to
confirm.
Not all the alarm limits can be deactivated!