Check the water trap
– Regularly check the water trap on the left-hand side
of the Cicero EM. When the level comes close to the
»Full«
mark, remove the jar by pulling it down and
empty it.
– Fit the jar firmly back in place!
If the permitted maximum level is exceeded, the
measured gas sampling system is switched off
automatically.
The advisory message »
WATER TRAP? !
« is displayed on
the system screen.
CO
2
, O
2
, breathing gas and N
2
O measurements will
then no longer be possible!
Remedy:
Drain the water trap as described.
Check water separators
– Water separators must be fitted at the lowest point in
the hose and hang downwards.
– They must be checked regularly and drained if
necessary.
Note hygiene regulations -
risk of infection!
The hose system remains sealed.
– Container must be replaced securely.
Anaesthetic gas scavenging line (AGS)
– Check the transfer hose.
Condensate may have collected in the hose – drain if
necessary.
40
Anaesthesia ventilation
Condensation