Message, cause and remedy
Messages on the system screen
(In alphabetical order according to priority)
Messages
Cause
Remedy
MEAN IBP1
<
!!
MEAN IBP2
<
!!
The average invasively measured blood pressure of
the displayed channel (1 or 2) is below the set
lower alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
MEAN IBP1
>
!!
MEAN IBP2
>
!!
The average invasively measured blood pressure of
the displayed channel (1 or 2) is below the set
lower alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
PRESS LIMIT !!, (!)
Ventilator operating with pressure limitation.
Change in lung compliance? Tube buckled?
Microbial filter in inspiratory line blocked?
Check tube/filter.
Increase Pmax or decrease V
T
if
necessary.
If pressure-limited ventilation is
intentional: press the »
config
«,
softkey and select »
alarms
« ; in
the alarms menu, switch »
Pmax
«
to »
off
« , to reduce the warning
message to a caution message
with no audible alarm.
PULSE IBP 1
>
!!
PULSE IBP 2
>
!!
Invasive measured pulse exceeds set alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
PULS SPO2
>
!!
Pulse rate exceeds upper alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
S-T MV
>
!!!
Magnitude of S-T segment exceeds upper alarm
limit.
Check patient's condition!
Correct alarm limit if necessary.
SPO2
>
!!
Oxygen saturation exceeds set upper alarm limit.
Check O
2
concentration in fresh
gas.
Check ventilator.
SYS IBP1
<
!!
SYS IBP2
<
!!
Measured invasive systolic pressure of the channel
shown (1 or 2) below set alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
SYS IBP1
>
!!
SYS IBP2
>
!!
Measured invasive systolic pressure of the channel
shown (1 or 2) exceeds set alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
SYS NIBP
<
!!
Measured non-invasive systolic pressure below set
alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
SYS NIBP
>
!!
Measured non-invasive systolic pressure exceeds
set alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
TEMP 1
<
!!
TEMP 2
<
!!
Measured body temperature of the channel shown
(1 or 2) below set alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
TEMP 1
>
!!
TEMP 2
>
!!
Measured body temperature of the channel shown
(1 or 2) exceeds set alarm limit.
Check patient's condition!
Correct alarm limit if necessary.