System screen functions
Configuring the system screen in »Standby«
SEF / SMF / EMG
SMF / d-Ratio / EMG
SEF / d-Ratio / EMG
SEF / d-Ratio / d / t / a / b
L.SEF / R.SEF / EMG
L.SEF / R.SEF / SMF
SMF
Hz
d-Ratio
0.5
10.5
EMG
SEF
90
Hz
SMF
12.5 10.5
EMG
SEF
90
Hz
d-Ratio
12.5
0.5
EMG
SEF
90
Hz
d-Ratio
12.5
0.5
d/q/a/b
L.SEF
90
Hz R.SEF
90
12.5 10.5
EMG
L.SEF
90
Hz R.SEF
90
12.5 10.5
L.SMF
9.5
R.SMF
8.5
V
T
0.60 L
1.25
Volumeter
4.3 L
30 s
12.5
Total Ampl. delta
47
µ
V 34
BSR
Amplitude
pEEG modules
pEEG modules are available for
selection in the parameter menu
(see page 52).
Please observe the separate Instructions
for Use!
Volumeter module
This module comprises a breath volumeter (upper bar
graph) and a minute volumeter (lower bar graph). The
length of the bar represents the maximum measured
value in accordance with the mode selected:
Mode
Breath V
T
MV
Neonates
0.25 L
2.5 L/min
Adults
1.25 L
12.5 L/min
Breath volumeter:
– The expiratory volume of each breath (V
T
) is shown
graphically by the upper bar and numerically in large
numerals abore.
Minute volumeter:
– The minute volumeter in the lower bar graph runs for
60 seconds, during which it adds up the tidal
volumes.
The time expired is shown in seconds together with the
value in large numerals. The individual breaths are shown
as segments in the bar graph. The volumeter stops
automatically after 60 seconds and beeps. The measured
values are displayed for 4 minutes and then deleted.
To start the volumeter:
●
Press the rotary control. If it is pressed again during
the 60 seconds, the values are deleted and the
volumeter restarted.