80
System screen functions
Alarm concept
Alarm modes
For adults / neonates:
The device stores two different sets of alarm limits, one
for adults and one for neonates.
The standard alarm limits for neonates and adults can
be configured in the Standby configuration menu.
The required mode (»
adult
« or »
neon.
«) is activated
under »
mode
« (see page 49).
The current mode is indicated in the status field at the top
left of the screen.
Activating default alarm settings
The alarm limits are reset to the preconfigured standard
alarm limits via the function »
default
« in the configuration
menu. The standard alarm limits are automatically active
after starting from »
Standby
« mode. The standard alarm
limits are automatically active after starting the screen
from »
Standby
« mode and after changing mode between
»
adult
« / »
neon.
«.
CO
2
alarm on/off
In »
MAN/SPONT
« monitoring mode, the alarm limits for
inspiratory and expiratory CO
2
can also be deactivated
by the »
CO
2
alarm on/off
« key.
In this case there will be no CO
2
monitoring!
Press the key again to reactivate the CO
2
alarms.
When changing over to
»IPPV«
monitoring mode or
»SIMV«
with a frequency of more than 6 ventilation
strokes per minute, the CO
2
alarms are automatically
reactivated.
Changing to
»HLM«
mode or
»SIMV«
with a frequency up
to 6 strokes/min. does not affect the deactivation.
When the CO
2
alarm is on and the ventilator is in IPPV
mode, the inspiratory CO
2
concentration in the breathing
gas is monitored with a fixed maximum limit of 5 mmHg!
The inspiratory CO
2
concentration can be set by the user
in the range from 0 to 10 mmHg.
Adaption to the ventilation mode
The alarm limits are switched over automatically with the
ventilator.
The alarm mode »
IPPV alarm limits
« is automatically
activated on the system screen if the ventilation
frequency for »
IPPV
« and »
SIMV
« is greater than
6 breaths per minute.und blinkt in den ersten fünf
Sekunden.
Check system configuration. Please confirm !
IPPV alarm limits
11:59
00:00
Anesth.Agent
selection
man. auto
gas
Isoflurane
The anesthetic
agent is automati-
cally selected by
agent detection.
Settings
mode
adult neon.
alarm sound
1 2 3 4 5 6 7 8 9
ECG pulse tone
0 1 2 3 4 5 6 7 8 9
ECG pulse tone
SpO
2
modulated on off
curve speed 12,5 25 50
mm/s
colours
call standard
Alarms
default
CO
2
on off
SpO
2
on off
parameter-
box
on off
HLM
on off
Pmax
on off
AGas Mix
on off
Calibration
iBP 1
iBP 2
more
?
?
06-12-98
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
1mV
iBP1
160
0
0
CO
2
40
30
33
6.0
freq 10 V
T
0.60
65
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat.alarm.
timer
start
para-
meter
config.
alarm
info
screen
config.
limits
auto set
vent. al.
2. ECG
II
pulse def 4 ST + 0.05
pulse 65
99
V
T
0.60
volumeter
4.3
27s
12
1.0
NiBP mmHg
Sys Dia
123/ 88
15 min
(102)
etCO
2
mmHg
MV L/min
HR/Pulse ECG
1/min
IBP1 ART mmHg
122
/
89
SpO
2
%
06-12-98
14029278
14229278